Changing the vacuum connector at, Pressure transducer vsk 3000 – VACUUBRAND DCP 3000 User Manual
Page 28
Installing the hose nozzle:
Unscrew the flange (using a 17 mm open-end wrench, if necessary) to
expose the compression fitting (1). Slip the compression nut, and then
the ferrule, onto the smooth end of the supplied hose nozzle (2). Slide the
smooth end of the hose nozzle onto the compression fitting on the VSK
3000 gauge head, and tighten the compression nut firmly finger-tight (3).
(1)
(3)
(2)
Installing the PTFE-tubing connection (PTFE-tubing ID: 8mm, OD:
19mm):
Unscrew the flange (using a 17 mm open-end wrench, if necessary) to
expose the compression fitting (1). Slip the compression nut, and then the
ferrule, onto the PTFE-tubing (2). Slide the PTFE-tubing onto the compres-
sion fitting on the VSK 3000 gauge head, and tighten the compression nut
firmly finger-tight (3).
Changing the vacuum connector at pressure transducer
VSK 3000
(1)
(3)
(2)