Measuring differential pressure, Automatic switching between pressure transducers, Vsk 3000 and vsp 3000 – VACUUBRAND DCP 3000 User Manual

Page 38

Advertising
background image

page 38 of 64

2nd possibility:
+

Switching during pressure indication or graphic display

Press the selection knob and keep pressed.

Turn the selection knob while still pressed until the desired pressure

transducer is indicated (see VACUU•BUS address in the status line).

Release the selection knob.

+

The selected pressure transducer is displayed in the status line as the

active pressure transducer.

Automatic switching between pressure transducers VSK

3000 and VSP 3000

If both a pressure transducer VSK 3000 and VSP 3000 are connected to

the vacuum gauge DCP 3000, the DCP 3000 automatically displays the

readings of the pressure transducer more suitable in the current pressure

range.

This requires both pressure transducers to be configured to corresponding

VACUU•BUS addresses (i.e. VSK x and VSP x). If the two VACUU•BUS

addresses do not correspond (i.e. VSK x and VSP y), an automatic switch-

ing between pressure transducers is not possible. It then is necessary to

reconfigure the address of one pressure transducer in order to enable

automatic switching (see section ”Reconfiguring the pressure transduc-

er” below).

Switchover points between pressure transducers VSK 3000 and

VSP 3000:

In case of decreasing pressure: Switching from VSK 3000 to VSP 3000

at a pressure of 0.7 Torr (1 mbar)

In case of increasing pressure: Switching from VSP 3000 to VSK 3000 at

a pressure of 3.7 Torr (5 mbar)

The active pressure transducer is shown in the status line. Furthermore,

a VSK 3000 is always displayed in decimal notation, a VSP 3000 in ex-

ponential notation.

Measuring differential pressure

A differential pressure between two pressure transducers can only be