Troubleshooting – VACUUBRAND MPT 200 User Manual

Page 13

Advertising
background image

page 13 of 22

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999226 // 05/05/2014

Troubleshooting

Fault

Possible cause

Remedy

❑ No display.

➨ Power supply (wall plug) not

plugged in, no power avail-

able?

✔ Plug in power supply, con-

nect cable of power supply

(wall plug) to DCP 3000.

Switch on device. Check

mains fuse.

➨ DCP 3000 switched off?

✔ Switch device on.

➨ Power supply (wall plug) not

connected to DCP 3000?

✔ Connect cable of power sup-

ply (wall plug) to DCP 3000.

➨ Other causes (device defec-

tive)?

✔ Contact local distributor.

❑ Malfunction of gauge

head (warning triangle

and display are flashing).

➨ MPT 100 / MPT 200 not con-

nected?

✔ Plug the MPT 100 / MPT 200

into the socket of the DCP

3000.

➨ MPT 100 / MPT 200 or con-

necting cable defective?

✔ Replace MPT 100 / MPT 200

or connecting cable if neces-

sary.

➨ Cold cathode not started?

✔ Wait for starting of the cold

cathode and allow for the

stabilization time. Clean the

gauge head, if necessary.

➨ Address of the MPT 100 /

MPT 200 changed?

✔ Remove rubber plug from the

address switch and set the

address to «1». Insert rubber

plug.

❑ The pressure reading

is too high in the high

vacuum range.

➨ Vacuum gauge head not cor-

rectly adjusted, cold cathode

does not start?

✔ Adjust vacuum gauge head

MPT 100 / MPT 200 under

vacuum.

➨ Vacuum gauge head contami-

nated?

✔ Clean the gauge head.

➨ Type of gas not considered?

✔ Comply with correction fac-

tors.

❑ The pressure reading is

too low at atmospheric

pressure.

➨ Vacuum gauge head not cor-

rectly adjusted?

✔ Readjust vacuum gauge

head correctly at atmospheric

pressure.

➨ Vacuum gauge head contami-

nated?

✔ Clean the gauge head.

➨ Deviation within the range of

the measurement uncertainty?

✔ Use pressure transducer

VSK 3000 (see accessories).

❑ Displayed pressure value

is fluctuating.

➨ Pressure fluctuations due

to the set-up of the vacuum

system?

✔ No measuring fault. Check

and change set-up of the

vacuum system if necessary.

➨ Plug of the cable to the vacu-

um gauge head not correctly

plugged in?

✔ Plug the cable correctly into

the socket at the rear of the

DCP 3000.