Type

MPT 200

Measuring principle

combined vacuum gauge head comprising two

components: thermal conductivity according to Pi-

rani and cold cathode tube; dependent on gas type

Measuring range (absolute)

1*10

3

mbar - 5*10

-9

mbar

(7.5*10

2

Torr - 3.7*10

-9

Torr)

Repeatability

(in percent of displayed value)

1*10

1

mbar - 1*10

-2

mbar:

(7.5*10

0

Torr - 7.5*10

-3

Torr):

1*10

-2

mbar - 1*10

-8

mbar:

(7.5*10

-3

Torr - 7.5*10

-9

Torr):

±2%

±7%

Measurement uncertainty

(in percent of displayed value

at constant temperature)

1*10

3

mbar - 1*10

2

mbar:

(7.5*10

2

Torr - 7.5*10

1

Torr):

1*10

2

mbar - 1*10

1

mbar:

(7.5*10

1

Torr - 7.5*10

0

Torr):

1*10

1

mbar - 2*10

-3

mbar:

(7.5*10

0

Torr - 1.5*10

-3

Torr):

2*10

-3

mbar - 1*10

-8

mbar:

(1.5*10

-3

Torr - 7.5*10

-9

Torr):

±50%

±30%

±10%

±25%

Maximum permissible pressure

at vacuum gauge head (absolute)

1.5 bar (1125 Torr)

Measuring cycle

10 ms

Maximum permissible temperature of gaseous

media at measurement connection

continuous operation: 60°C,

for short periods (less than 5 minutes) up to 80°C

Maximum permissible temperature during

bake-out

max. 180°C

(without electronics and magnetic unit)

Ambient temperature range (operation)

+5°C to +60°C

Ambient temperature range (storage)

-40°C to +65°C

Permissible relative atmospheric moisture

during operation (no condensation)

30% to 85%

Maximum permissible installation altitude above

mean sea level

2000 m

Range of supply voltage

(via VACUU•BUS cable)

24 VDC

Max. power consumption (supply)

3 W

Degree of protection according to IEC 529

IP 54

Weight

555 g

Dimensions L x W x H

138 mm x 66 mm x 53 mm

Vacuum connection

small flange KF DN 25

Cable length

approx. 2.0 m