Notes on operation

For operation it is necessary to install valves and/or vacuum pumps. Without those components,

the controller will work only as a vacuum measurement device. As such the controller will switch

a possibly connected coolant valve.

- Inside a vacuum system where evaporation occurs, the

vacuum is not uniform. This effects the value of the meas-

ured pressure and such the controlling. Therefore carefully

choose the position where to connect the pressure trans-

ducer.

- Condensate and deposits at the pressure transducer fal-

sify the measurement result.

- If residues occur or when working with aggressive or con-

densable substances, install a gas washing bottle in front

of the pressure transducer.

-

Position controller in such a way, that condensate can

not flow into the pressure transducer.

- If necessary, clean the pressure transducer, see section

„Cleaning the pressure transducer“, pg. 36.

Setting of interface parameters, see

„Interface parameters“, pg. 26.

Presettings at controller, see „Menu structure of controller“, pg. 18.

Operating the controller, see

„Working with the controller“, pg. 21.

Note: Texts written in Courier font mirror the display of the LCD of the controller.

Connecting an external pressure transducer VSK 3000 will automatically replace the internal pressure

transducer as active sensor. The display will show the vacuum measured by the VSK 3000, the vacuum

connection of the controller is rendered inoperable.

Attention: At pressures above approximately 1100 mbar the pressure reading

becomes incorrect due to saturation of the pressure transducer.

Release pressure immediately! Risk of bursting!

At pressures above 1060 mbar the

display flashes.

Maximum permissible pressure: 1.5 bar absolute.

When switching on the controller VNC 2 for the very first time, a menu to select the language of the

controller menu is displayed. Select the desired language (”Deutsch”, ”English”, ”Français”) using the

keys t and s, and press ”Enter” to confirm. Then select the pressure unit (”mbar”, ”Torr” or” hPa”) in

the same way.

vacuum connection to application

to pump