Read commands compatible to ”cvc 3000 – VACUUBRAND VNC 2 E User Manual

Page 30

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999199 / 28/11/2012

page 30 of 41

0: remote operation off

1: remote operation on

1.......9: sensor quantity

1.......9: number of active sensor

0: remote module not connected

1: remote module connected

0: level sensor not connected

1: level sensor connected

0: fault indicator not connected

1: fault indicator connected

0: venting valve not connected

1: venting valve connected

0: coolant valve not connected

1: coolant valve connected

0: in-line valve not connected

1: in-line valve connected

0: VMS not connected

1: VMS connected

0: NT VARIO pump not connected

1: NT VARIO pump connected

0: acoustic signal off

1: acoustic signal on

0: autostart off

1: autostart on

0: pressure unit mbar

1: pressure unit Torr

2: pressure unit hPa

0: language German

1: language English

2: language French

0: VACUULAN

1: continuous pumping

2: vacuum control

4: RC 5/6 mangement

Description

unit according to preselections

1-100% or HI

process runtime

VACUU•LAN: runtime of pump since

last switching of the pump

Response

XXXX.X mbar/Torr/hPa

XXX%

XX:XX hh:mm

xxx/min

XX:XX hh:mm

XXXXXXXXXXXXXXXX

Function

actual vacuum

actual speed

time

LAN

pressure increase

LAN process time

preselected

at the controller

Command

IN_PV_1

IN_PV_2

IN_PV_3

IN_PV_4

IN_PV_5

IN_CFG

Read commands compatible to ”CVC 3000”