Avoid high heat supply (e.g., due to hot process gases).

Position device and vacuum connection lines so that no condensate can flow to-

wards the pressure transducer.

Allow the equipment to equilibrate to ambient temperature if you bring it from cold

environment into a room prior to operation. Notice if there is water condensation on

cold surfaces.

Comply with all applicable and relevant safety requirements (regulations and guide-

lines).

Implement the required actions and adopt suitable safety measures.

Ambient conditions

Ensure that installation is in compliance with the degree of protection, see “Technical

data“.

Operating conditions

This device is not approved for operation in potentially explosive atmospheres.

Do not operate the device in potentially explosive atmospheres.

Devices without the ”

`

” mark on the rating plate are not approved for op-

eration with dangerous or explosive gases or with potentially explosive or

inflammable substances. Do not operate the device with dangerous or ex-

plosive gases or with potentially explosive or inflammable substances.

Devices bearing the ”

`

” mark on their rating plates are approved for op-

eration with potentially explosive atmospheres according to their classifica-

tion

II 3G IIC T3 X according to ATEX, but they are not approved for operation

in potentially explosive atmospheres (see section

`

Important information:

Equipment marking (ATEX)”)

.

• Ensure that the materials of the equipment’s wetted parts are compatible with the

substances in the vacuum system, see section ”Technical data”.

Safety during operation

Adopt suitable measures to prevent the release of dangerous, toxic, explosive,

corrosive, noxious or polluting fluids, vapours and gases.

Prevent any part of the human body from coming in contact with the vacuum.

VSK 3000:

Attention: At pressures above 1060 mbar the pressure transducer is saturated,

the pressure measurement becomes erroneous.

Release pressure immediate-

ly. Risk of bursting!

Maximum permitted pressure: 1.5 bar (absolute).

VSP 3000:

Attention: Maximum pressure output: 1*10

3

mbar. Pressure values above 1000

mbar can not be read out!

Danger of unnoticed overpressure! Risk of bursting!

Maximum permitted pressure: 1.5 bar (absolute).