Use and operation

Notes on operation

Gas

Correction factor cold cathode

Correction factor Pirani

Air / N

2

-

1.00

CO

2

0.74

0.89

H

2

2.39

0.58

He

5.93

1.02

Ne

3.50

-

Ar

0.80

1.59

Kr

0.60

-

Xe

0.41

-

The vacuum gauge head MPT 100 consists of two components: one cold cathode sensor and one Pirani

sensor. The pressure measurement is dependent on the gas type. The vacuum gauge head is factory-

set for operation with air / N

2

.

Use the following correction factors when working with different gases in the pressure range below 0.1

mbar:

Connecting and starting operation

Vacuum connection:

Remove the protective cover and keep.

Connect the vacuum gauge head via small flange to the vacuum application.

The vacuum gauge head may be mounted in any orientation, though an orienta-

tion is to be preferred in which condensate cannot get into the measuring cham-

ber.

Avoid contamination (oil/oil mist) of the measurement chamber when generating the vacuum with an

oil-filled vacuum pump.

Do not mount the vacuum gauge head directly at the oil-sealed pump. The diameter of the pipelines

should be as large as possible.

Inside a vacuum system where evaporation occurs (e.g. rotary evaporator) the vacuum is not uni-

form. Therefore choose carefully the position where to connect the vacuum gauge head.

Condensate and deposits in the measuring chamber falsify the measurement result.

Soiling and damage effect adversely the function of the vacuum gauge head.

If residues occur or when working with aggressive or condensable substances, install a gas washing

bottle in front of the vacuum gauge head MPT 100.

Electrical connection / VACUU•BUS

Voltage supply and readout of the vacuum gauge head via VACUU•BUS line by the vacuum gauge

DCP 3000 using VACUU•BUS protocol. Maximum cable length inside buildings: 30m. Extension cable

VACUU•BUS 2m: order-no. 612552.