Hanna Instruments HI 981400 User Manual

Hanna Instruments Sensors

Advertising
background image

Instruction Manual

HI 981400

pH Indicator

with Alarm Signal for

HACCP Control

5.8

FOOD STABILITY
p H I N D I C ATO R

HACCP

pH

pH4

pH7

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

WARRANTY

The HI 981400 pH Indicator is warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. This warranty is lim-
ited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

w w w . h a n n a i n s t . c o m

OPERATIONAL GUIDE

pH ELECTRODE

• Any combination pH electrode provided with a BNC

connector can be attached to the controller.
See our General Catalog for a complete range of Hanna
pH electrodes.

• Do not be alarmed if white crystals appear around the

electrode protective cap. This is normal with pH elec-
trodes and they dissolve when rinsed with water.

• When not in use, rinse the electrode with water and

store it with a few drops of storage (HI 70300) solution
in the protective cap.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STOR-
AGE PURPOSES.

• If the electrode has been left dry, soak the tip in a

storage (HI 70300) or pH 7 (HI 7007) solution for at
least one hour to reactivate it.

• To minimize clogging and provide longer life for the pH

electrode, it is recommended to clean it monthly.
Immerse the tip of the electrode in the HI 7061 solution
for half an hour, then rinse it with tap water.

Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this manual before using the meter.
If you need additional technical information, do not hesitate
to e-mail us at [email protected].
This instrument is in compliance with the

directives.

Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:

• Calibration screwdriver
• pH 4.01 and 7.01 buffer solutions (20 mL each)
• 12 Vdc power adapter and instructions

Note: Conserve all packing material until the instrument

has been observed to function correctly. Any defective
item must be returned in its original packing.

HI 981400 is a pH indicator specially designed for pH
control in foodstuffs.
The pH value is one of the most important indicators of food
quality and safety. pH of raw ingredients, such as milk and
meat, is measured to ensure the quality standards of
products. pH is also monitored at different stages of food
processing to guarantee safety and enhance quality.
HI 981400 is also provided with a flashing LED that
indicates if the measured pH exceeds the selected limit.
Hanna Instruments offers several pH electrodes specially
designed for the food industry, such as the FC 200B with
open junction, and the FC 230B with stainless steel blade
for direct meat measurement.
Thanks to the built-in hook, the meter can b easily hanged
right above the sample for continuous monitoring.
The meter can be calibrated at one or two points.
You no longer need to worry about battery changes because
the unit runs without interruption on 12 Vdc power supply.

FUNCTIONAL DESCRIPTION

1. Molded eye
2. Alarm LED
3. “pH 4” calibration trimmer
4. BNC connector for pH electrode
5. Liquid Crystal Display
6. “pH 7” calibration trimmer
7. Power supply connector
8. 12 Vdc power adapter (included)

5.8

FOOD STABILITY
p H

I N D I C ATO R

HACCP

pH

pH4

pH7

Any pH electrode ending in a

BNC connector can be attached

to the meter

1

2
3

4

6

5

7

8

SPECIFICATIONS

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy

(@ 25°C/77°F)

±0.2 pH

Typical EMC Deviation

±0.2 pH

Setpoint

4.6 to 7.5 pH

Alarm

LED blinks when

measurement exceeds the setpoint value

Calibration

Manual, 1 or 2 point, with

offset (pH 7) and slope (pH 4) trimmers

Power Supply

12 Vdc adapter (included)

Dimensions

86 x 110 x 43 mm (3.4 x 4.3 x 1.7’’)

Weight

150 g (5.3 oz.)

Advertising