5. CALIBRATION AND MAINTENANCE

BETA D-50

BATTERY

R

SHUNT

AMMETER

mV

EXTERNAL CALIBRATED

MILLIVOLTMETER

TEST

PORTS

Figure 5-2 Circuit diagram for ammeter calibration

A.

SWITCH OFF MAINS POWER

Turn off the AC on-off/reset power switch.

B.

TURN DOWN DISCHARGE CUR-

RENT

Repeatedly turn the Current Adjust knob fully counter-clockwise to set discharge
current to zero.

C.

CONNECT BATTERY

Connect the battery DC cable to the battery and ensure the connector is plugged
in completely.





















Page 22

BETA D-50 TESTER / ANALYZER - OPERATING MANUAL V2.2