VERIFYING METER READING (NOT CALIBRATION)

1. GENERAL:

The CA 1550 has been factory calibrated prior to shipment and

the unit can be checked for verification of readings of the voltmeter and

ammeter with a Fluke 77/AN multimeter. The ampere-hour meter and

ammeter must be checked with a load.

2. CHECKING THE VOLTMETER WITHOUT THE BATTERY CONNECTED:

a) On / Off switch OFF (10).

b) Charge / Discharge switch to Charge (11).

c) Charge Current Adjust Max. (5).

d) Insert external DC Voltmeter leads into the panel volts test points (7).

e) On / Off switch to On (10).

f) Verify that the panel meter and test meter such as fluke 77/AN on the DC

Volts scale read within + - 0.3 Volts.

g) Adjust Voltage to confirm that the readings coincide over a voltage range

e.g.:22-30 Volts.

h) Reset Voltage adjust as required when the check is completed e.g.: 28.5

Volts.

3. CHECKING VOLTMETER WITH THE BATTERY CONNECTED:

a) Perform as above, except start with the Charge Current Adjust (5) set to

minimum. Increase current to increase and check the voltage. (Do not exceed

any maximum current requirements).