Fluke 87 V User Manual

Page 16

Advertising
background image

80 Series V
Users Manual

8

Table 4. Pushbuttons

Button

Switch

Positon

Function

N

Selects capacitance

M

Selects temperature (87 only)

J

Selects ac low pass filter function (87 only)

P

Switches between dc and ac current

Q

Switches between dc and ac current

A

(Yellow)

Power-up

Disables automatic power-off feature (Meter normally powers off in 30 minutes).
The Meter reads

PoFF”

until

A

is released.

B

Any switch

position
Power-up

Starts recording of minimum and maximum values. Steps the display through MAX, MIN, AVG

(average), and present readings. Cancels MIN MAX (hold for 1 second)
Enables the Meter’s calibration mode and prompts for a password.
The Meter reads “

CAL

and enters calibration mode. See 80 Series V Service Information

.

C

Any switch

position

Switches between the ranges available for the selected function. To return to autoranging, hold

the button down for 1 second.

M

Power-up

Switches between

°

C and

°

F.

Enables the Meter’s smoothing feature. The Meter reads “

5___” until

C

is released.