80 Series V
Users Manual

14

When measuring voltage, the Meter acts approximately
like a 10 M

Ω

(10,000,000

Ω

) impedance in parallel with

the circuit. This loading effect can cause measurement
errors in high-impedance circuits. In most cases, the error
is negligible (0.1% or less) if the circuit impedance is
10 k

Ω

(10,000

Ω

) or less.

For better accuracy when measuring the dc offset of an ac
voltage, measure the ac voltage first. Note the ac voltage
range, then manually select a dc voltage range equal to or
higher than the ac range. This procedure improves the
accuracy of the dc measurement by ensuring that the
input protection circuits are not activated.

MIN MAX

RANGE

REL

Hz %

AutoHOLD

Peak MIN MAX

4½ DIGITS

1 Second

˚

C/

˚

F

OFF

mA

A

mV

V

V

A

A

mA

COM

V

400mA
FUSED

10A MAX

FUSED

A

TRUE RMS MULTIMETER

87 V

LO

LO

MIN MAX

RANGE

REL

Hz %

AutoHOLD

Peak MIN MAX

4½ DIGITS

1 Second

˚

C/

˚

F

OFF

mA

A

mV

V

V

A

A

mA

COM

V

400mA
FUSED

10A MAX

FUSED

A

TRUE RMS MULTIMETER

87 V

LO

LO

Switch Box

V

+

AC Voltage

DC Voltage

V

mV

aom2f.eps

Figure 2. Measuring AC and DC Voltage

Advertising