Detailed specifications – Fluke 87 V User Manual

Page 52

Advertising
background image

80 Series V
Users Manual

44

Detailed Specifications

For all detailed specifications:
Accuracy is given as

±

([% of reading] + [number of least significant digits]) at 18° C to 28° C, with relative humidity up to

90 %, for a period of one year after calibration.
For Model 87 in the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10. AC conversions are ac-

coupled and valid from 3 % to 100 % of range. Model 87 is true rms responding. AC crest factor can be up to 3 at full scale, 6

at half scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.

Table 10. Model 87 AC Voltage Function Specifications

Function Range Resolution

Accuracy

45 – 65 Hz 30 – 200 Hz

200 – 440 Hz

440 Hz - 1 kHz

1 - 5 kHz

5 - 20 kHz

[1]

600.0 mV

0.1 mV

±

(0.7 % + 4)

6.000 V

0.001 V

60.00 V

0.01 V

±

(2.0 % + 4)

±

(2.0 % + 20)

600.0 V

0.1 V

±

(2.0 % + 4)

[3]

unspecified

1000 V

1 V

±

(0.7 % + 2)

±

(1.0 % + 4)

unspecified unspecified

K

[2,4]

Low pass filter

Same as
45-65 Hz

±

(1.0 % + 4)

+1 % + 4

-6 % - 4

[5]

unspecified unspecified unspecified

[1]

Below 10 % of range, add 12 counts.

[2]

The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between
1 and 30 counts. A 30 count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce
a much larger constant error in later measurements.

[3]

Frequency range: 1 kHz to 2.5 kHz.

[4]

A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.

[5]

Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.