MIN MAX AVG Recording Mode

The MIN MAX AVG recording mode captures the minimum and

maximum input values, and calculates a running average of all

readings. When a new high or low is detected, the Meter beeps.

Note

For DC functions, accuracy is the specified accuracy of the

measurement function

±

12 counts for changes longer than

350 ms in duration.
For AC functions, accuracy is the specified accuracy of the

measurement function

±

40 counts for changes longer than

900 ms in duration.

To use MIN MAX AVG recording:
⇒ Make sure that the Meter is in the desired measurement

function and range. (Autoranging is disabled in the MIN MAX

AVG mode.)

⇒ Press

MIN MAX to activate MIN MAX AVG mode.

and MAX light, and the highest reading detected

since entering MIN MAX AVG is displayed.

⇒ Press

MIN MAX to step through the low (MIN), average

(AVG), and present readings.

⇒ To pause MIN MAX AVG recording without erasing stored

values, press HOLD. h is displayed.
To resume MIN MAX AVG recording, press HOLD again.
h

turns off.

⇒ To exit and erase stored readings, press MIN MAX for 1

second or turn the rotary switch.

Display HOLD and AutoHOLD Modes

XW

Warning

To avoid electric shock, do not use the Display HOLD or

AutoHOLD mode to determine if a circuit is live.

Unstable or noisy readings will not be captured.

In the Display HOLD mode, the Meter holds the reading on the

display.
In the AutoHOLD mode, the Meter holds the reading on the display

until it detects a new stable reading. Then the Meter beeps and

displays the new reading.
⇒ Press HOLD to activate Display HOLD. h lights.
⇒ Press HOLD again to activate AutoHOLD. Ah lights.
⇒ Press HOLD again to resume normal operation.
To resume normal operation at any time, press HOLD for 1

second or turn the rotary switch.

YELLOW Button

Press the YELLOW

button to select alternate measurement

functions on a rotary switch setting, e.g., to select DC mA, DC A,

Hz, temperature (Model 179 only), capacitance, diode test.

Display Backlight (Model 177 and 179 Only)

Press

S

to toggle the backlight on and off. The backlight

automatically turns off after 2 minutes.