Auto ramping the ma output – Fluke 789 User Manual

Page 34

Advertising
background image

789
Users Manual

26

Table 9. mA Stepping Pushbuttons

Pushbutton Adjustment

X

M

% STEP

Adjusts up to the next higher 25 %

step

% STEP

G

W

Adjusts down to the next lower 25

% step

Span Check

Sets to 100 % value

Span Check

Sets to 0 % value

Table 10. mA Step Values

Value (for each span setting)

Step

4 to 20 mA

0 to 20 mA

0 %

4.000 mA

0.000 mA

25 %

8.000 mA

5.000 mA

50 %

12.000 mA

10.000 mA

75 %

16.000 mA

15.000 mA

100 %

20.000 mA

20.000 mA

125 %

24.000 mA

120 %

24.000

mA

Auto Ramping the mA Output

Auto ramping gives the ability to continuously apply a
varying current stimulus from the meter to a transmitter,
while hands remain free to test the response of the
transmitter. Select either sourcing or simulating by
choosing the SOURCE or SIMULATE jacks.