Calibrating an I/P Device

43

Calibrating an I/P Device

The following test allows you to calibrate a device that
controls pressure. Proceed as follows:

1. Connect the test leads to the instrument under test as

shown in Figure 21. The connections simulate a
current-to-pressure transmitter and measures the
corresponding output pressure.

2. Press

A

(upper display).

3. Press

V

for sourcing current (lower display).

4. If necessary, press

M

for SOURCE mode.

5. Enter the desired current by pressing

X

and

W

keys. Press

Y

and

Z

to select different digits.