Remote control commands – Fluke 725 User Manual

Page 56

Advertising
background image

725
Users Manual

46

Remote Control Commands

You can control the calibrator remotely from a PC running
a terminal emulator program. The remote control
commands give access to all capabilities of the calibrator
with the exception of pressure measurement. See Table
8A-8C for the remote commands and explanations.

The Fluke 700SC Serial Interface Cable Assembly (PN
667425) plugs into the pressure module connector and

terminates in a DB-9 connector which plugs directly into a
PC serial port. A DB-9 to DB-25 adapter is required to
connect to a PC.

The remote control interface on the 725 is activated by
turning the calibrator off, then turning it on again while
depressing the

M

key. The calibrator will initialize with its

remote port enabled. The terminal emulator connected to
the calibrator should be set up for: 9600 baud, no parity,
8 data bits, and 1 stop bit.

Table 8A. Remote Control Upper Display

Serial Input

Description

j

mA measurement

L

mA Loop Power

E

Volts measurement

B

Single broadcast of most recent upper display value and units

Table 8B. Remote Control Lower Display

Serial Input

Description

A

mA measurement

a

mA source

I

mA 2W Sim

V

Volts measurement

v

Volts source

M

mV measurement

m

mV source

K

KHz measurement