Frequency source, Temperature, thermocouples – Fluke 725 User Manual

Page 68

Advertising
background image

725
Users Manual

58

Frequency Source


Range


Resolution

Accuracy

(% of output

frequency)

2.0 to 1000.0 CPM

0.1 CPM

±

0.05 %

1 to 1000 Hz

1 Hz

±

0.05 %

1.0 to 10.0 kHz

0.1 kHz

±

0.25 %

Waveform: 5 V p-p squarewave, -0.1 V offset

Temperature, Thermocouples

Type

Range

Measure and Source

Accuracies (ITS-90)

J

-200 to 0

°

C

0 to 1200

°

C

1.0

°

C

0.7

°

C

K

-200 to 0

°

C

0 to 1370

°

C

1.2

°

C

0.8

°

C

T

-200 to 0

°

C

0 to 400

°

C

1.2

°

C

0.8

°

C

E

-200 to 0

°

C

0 to 950

°

C

0.9

°

C

0.7

°

C

R

-20 to 0

°

C

0 to 500

°

C

500 to 1750

°

C

2.5

°

C

1.8

°

C

1.4

°

C

S

-20 to 0

°

C

0 to 500

°

C

500 to 1750

°

C

2.5

°

C

1.8

°

C

1.5

°

C

B

600 to 800

°

C

800 to 1000

°

C

1000 to 1800

°

C

2.2

°

C

1.8

°

C

1.4

°

C

L

-200 to 0

°

C

0 to 900

°

C

0.85

°

C

0.7

°

C

U

-200 to 0

°

C

0 to 400

°

C

1.1

°

C

0.75

°

C

N

-200 to 0

°

C

0 to 1300

°

C

1.5

°

C

0.9

°

C

XK

-200 to -100

°

C

-100 to 800

°

C

0.5

°

C

0.6

°

C

BP

0 to 800

°

C

800 to 2500

°

C

1.2

°

C

2.5

°

C

Resolution:
J, K, T, E, L, N, U, XK, BP:

0.1

°

C, 0.1

°

F

B, R, S:

1

°

C, 1

°

F

Advertising