Fluke 77 User Manual

Page 16

Advertising
background image

Introduction and Specifications

Specifications

1

1-7

Overload protection for all functions and ranges: Rated voltage.

Function

Input Impedance (Nominal)

L,

m

L, K

>10 M

, <50 pF

Common Mode Rejection
Ratio (1 k

Unbalanced)

Normal Mode Rejection

L,

m

L

>120 dB at dc, 50 Hz, or 60 Hz

>60 dB at 50 Hz or 60 Hz

K

>60 dB dc to 60 Hz

Open Circuit Test Voltage

Full Scale Voltage

To 3.2 M

32 M

e

<3.1 V dc
<2.8 V dc (typical)

<440 mV dc
<420 mV dc
(typical)

<1.4 V dc
<1.3 V dc
(typical)

Open Circuit Test Voltage

Full Scale Voltage

G R

<3.1 V dc

2.0 V dc

Short Circuit Current (typical)

e

300

µ

A

G R

400

µ

A

G

V

F

Current (typical)

0.0 V
0.6 V
1.2 V
2.0 V

0.4 mA
0.3 mA
0.2 mA
0.1 mA

Advertising