Ac voltage test, Resistance test, 8. ac voltage test – Fluke 77 User Manual

Page 33

Advertising
background image

77 Series III

Service Manual

3-8

Table 3-2. DC Voltage Test

Step

DC Input Voltage

Display Reading

1

+2.7V

2.691 to 2.709V DC

2

+27V

26.91 to 27.09V DC

3

+270V

269.1 to 270.9V DC

4

+1000V

995 to 1005V DC

5*

+300 mV

299.0 to 301.0V DC

*300 mV function only

3-8. AC Voltage Test

Warning

Connect the ground/common/low side of the AC
calibrator to COM on the UUT.

1.

Set the UUT function switch to VAC, and connect the AC Voltage
Calibrator to the

z and COM input terminals.

2.

Set the AC Voltage Calibrator for the output given in Table 3-3,
verify that the UUT display reading is within the limits shown.

Note

When the input is open in the VAC function, it is normal for
the meter to read some counts on the display. This is due to
ac pickup in the ac amplifier when the ac amplifier when the
ac amplifier is unterminated.

Table 3-3. AC Voltage Test

Input

Step

Voltage

Frequency

Display Reading

1

2.7V

100 Hz

2.644 to 2.756V AC

2

2.7V

500 Hz

2.644 to 2.756V AC

3

1000V

100 Hz

978 to 1022V AC

4

1000V

1000 Hz

978 to 1022V AC

3-9. Resistance

Test

1.

Select the ohms function on the UUT.

2.

Connect the Ohms Calibrator or Decade Resistor to the

z and

COM input terminals of the UUT.