Fluke 189 User Manual

Page 91

Advertising
background image

Specifications

Detailed Accuracy Specifications

7

7-7

Accuracy

Accuracy Dual Display AC or AC+DC

3

Function

Range

Resolution

DC

20 - 45 Hz

45 Hz - 1 kHz

1 kHz- 20 kHz

DC mV

50.000 mV

0.001 mV

0.1% + 20

6 % + 40

500.00 mV

0.01 mV

0.03 % + 2

3000.0 mV

0.1 mV

0.025 % + 5

2 % + 40

DC V

5.0000 V

0.0001 V

0.025 % + 10

2

50.000 V

0.001 V

0.03 % + 3

2

500.00 V

0.01 V

0.1 % + 2

2

Not specified

1000.0 V

0.1 V

0.1 % + 2

2

2 % + 80

0.5 % + 40

Not specified

DC

µ

A

500.00

µ

A

0.01

µ

A

0.25 % + 20

1 % + 20

1.0 % + 20

2 % + 40

5,000.0

µ

A

0.1

µ

A

0.25 % + 2

1 % + 10

0.75 % + 10

2 % + 40

DC mA

50.000 mA

0.001 mA

0.15 % + 10

1 % + 20

0.75 % + 20

2 % + 40

400.00 mA

0.01 mA

0.15 % + 2

1 % + 10

1 % + 10

3 % + 40

DC A

5.0000 A

0.0001 A

0.5 %+ 10

2 % + 20

2 % + 20

6 % + 40

10.000 A

1

0.001 A

0.5 %+ 2

1.5 % + 10

1.5 % + 10

5 % + 10

1. 10 A continuous up to 35 °C, less than 10 minutes 35 °C to 55 °C. 20 A overload for 30 seconds maximum.
2. 20 counts in dual display DC or AC+DC.
3. See AC conversions notes for AC mV and V

Advertising