DC Voltage Specification

Accuracy

AC over DC, DC over AC, AC + DC

[2]

Function Range Resolution

DC

[2]

20 to 45 Hz

45 Hz to 1 kHz

1 to 20 kHz

20 to 35 kHz

50 mV

[3]

0.001 mV

0.05 % + 20

[4]

1.5 % + 40

5 % + 40

DC mV

500 mV

0.01 mV

0.025 % + 2

[5]

1.5 % + 40

5 % + 40

5 V

0.0001 V

0.025 % + 2

1.5 % + 40

5 % + 40

50 V

0.001 V

0.025 % + 2

1.5 % + 40

5 % + 40

500 V

0.01 V

0.03 % + 2

Not Spec’d

Not Spec’d

DC V

[1]

1000 V

0.1 V

0.03 % + 2

2 % + 80

0.5 % + 80

Not Spec’d

Not Spec’d

L

[1]

1000 V

0.1 V

1 % + 20

Not Spec’d

Not Spec’d

Not Spec’d

Not Spec’d

[1]

Add 20 counts in dual display ac over dc, dc over ac or ac+dc.

[2]

AC+DC ranges are specified from 2 % to 140 % of range, except 1000 V is specified from 2 % to 100 % of range.

[3]

When using the relative mode (REL Q) to compensate for offsets.

[4]

Add 4 counts/10 mV AC in dual display ac over dc, dc over ac or ac + dc

[5]

Add 10 counts/100 mV AC in dual display ac over dc, dc over ac or ac + dc.