Input characteristics – Fluke 289 User Manual

Page 87

Advertising
background image

True-rms Digital Multimeters

Detailed Specifications

77

Input Characteristics

Function

Overload

Protection

[1]

Input Impedance

Common Mode
Rejection Ratio

(1 k

Ω

unbalance)

Normal Mode Rejection

L

1000 V

10 M

Ω

<100 pF >120 dB at dc, 50 Hz or 60 Hz

>60 dB at 50 Hz or 60 Hz

F

mV

1000 V

[2]

10 M

Ω

<100 pF >120 dB at dc, 50 Hz or 60 Hz

>60 dB at 50 Hz or 60 Hz

K

1000 V

10 M

Ω

<100 pF

(ac-coupled)

>60 dB, dc to 60 Hz

L

1000 V

3.2 k

Ω

<100 pF

(ac-coupled)

Not specified

Not specified

Full Scale Voltage

Typical Short Circuit Current

Function

Overload

Protection

[1]

Open Circuit
Test Voltage

To 500 k

Ω

>5 or 50 nS

500

Ω

5

k

Ω

50

k

Ω

500

k

Ω

5

M

Ω

50

M

Ω

e

1000 V

[2]

5 V dc

550 mV

<5 V

1 mA

100

μ

A 10

μ

A

1

μ

A 0.3

μ

A 0.3

μ

A

50e 1000

V

[2]

20 V decreasing
to 2.5 V

500 mV

10 mA

G

1000 V

[2]

5 V dc

3.1 V dc

1 mA

[1]

Input is limited to the product of a V rms sinewave times frequency of 2 x 10

7

V-Hz.

[2]

For circuits <0.5 A short circuit. 660V for high energy circuits.