Cascade Microtech InfinityQuad Probe Pad Layout Rules User Manual

Infinityquad probe: pad layout rules, Rule 100: pad size

Advertising
background image

PN 154-631-2

www.cascademicrotech.com

InfinityQuad Probe PAD Layout Rules • 1

Rule 100: Pad Size

Minimum pad sizes are shown in

Table 1

and

Table 2

.

The longer dimension on the rectangular pad represents the direction of probe scrub. To provide for probing ease and
parasitics low enough for many applications, use 100 x 100 µ pads for 150 µ pitch and above, unless there is a reason to
do otherwise. Increase pad size when using an autoprober to account for errors. The recommended amount of increase
depends on the station and the type of measurement. As a general rule, however, increase the pad size by 5 µ in both
length and width.

Table 1. Minimum pad size rule in 5 µ increments (units in µ x µ).

N

OTE

Recommended minimum pad sizes include error margins for automated measurements (on
a Cascade Microtech semi-automatic station) and over temperature probing (-40° to
+125°C).

# of Contacts

(include X)

75 µ

pitch

80 µ

pitch

100 µ

pitch

125 µ

pitch

150 µ

pitch

200 µ

pitch

250 µ

pitch

4

25 x 45

5
6
7
8
9

10

30 x 50

11
12
13
14
15
16

35 x 55

17
18
19

40 x 60

20
21
22
23

45 x 65

24
25

i