Percent of basal – Energizer 715 User Manual

Page 123

Advertising
background image

Optimizing pump therapy 101

Percent of basal

The temp basal type is dependent on your current basal rates. Percent temp basal is a percentage

increase or decrease of your current basal (0 - 200 percent limited to your maximum basal rate setting).

The maximum percent limit is based on the largest basal rate segment with your programmed temp

basal rate time.

For example: It is 6:00AM and your current basal rate is 1.50 U/H. You want to set a temp basal rate

of 130 percent for seven (7) hours. The maximum percent temp basal rate you can set is 125 percent.
Anything larger would make #2 segment exceed your maximum basal setting of 2.0 U/H.

If your current basal changes (i.e. from rate 1 to rate 2), your percent temp basal amount will also

change. The pump will deliver the percentage for the duration that you have set.

You cannot make changes to your normal basal rate while a percent temp basal is active. You must

either wait until the temp basal is finished or cancel the temp basal in order to reprogram your normal
basal rate setting(s).

Your current basal rates:

Your maximum basal rate setting: 2.0 U/H

Segment #1: 12:00A

1.50 U/H

Segment #2: 11:00A

1.60 U/H (largest)

Segment #3: 4:00P

1.30 U/H

Advertising
This manual is related to the following products:

515