Block feature

The block allows you to prevent (block) people from changing pump setting that should not be

working with the pump s settings. This feature is an important safety feature, especially if the pump
user requires someone else to maintain complete control of pump operation. The factory setting for
this feature is off. When block is on, the remote control is used to deliver a bolus and suspend the
pump. Direct pump programming is limited to suspend, block, and selftest. You can, however, still view
status-type screens (STATUS, BOLUS and PRIME HISTORY, BASAL REVIEW, DAILY TOTALS, etc.). Discuss
what uses and settings are best for you with your healthcare professional. (You can order the remote
control from Medtronic MiniMed.)

Turn block on

NOTE -

The vibrate alert-type is disabled when block is on.

1

Go to the BLOCK OPTION screen.

Main > Utilities > Block

2

Select

On

and press

ACT

. The HOME screen will

appear with an open circle. The Block option is now
on and the pump is in Special mode. Exit the menus.

Add

Delete

Review

Select

Add ID

and press

ACT

.

Select

Delete ID

and press

ACT

.

Select

Review ID

and press

ACT

.

Enter each of the six ID numbers.
Press

ACT

after each.

After you set the last number of
the ID, the screen will return to
the REMOTE ID MENU.

Select the remote ID that you
want to delete and press

ACT

.

The selected ID is now deleted.

The programmed IDs will show
in the REVIEW REMOTE ID
screen.

ADD REMOTE ID

_ _ _ _ _ _

(flashing)

DELETE REMOTE ID

2 222222
3 - - - - - -

1 111111

REMOTE ID MENU

Delete ID

Add ID

Review ID

REVIEW REMOTE ID

1 - - - - -
2 222222
3 - - - - -

BLOCK OPTION

Off

ACT

On

Advertising