Restoring pump settings, 124 chapter: 8 – Energizer 715 User Manual

Page 146

Advertising
background image

124 Chapter: 8

3

If this is the first time you have saved pump settings, go to step

4

.

If you have previously saved pump settings, a message displays,
indicating the date of your last pump settings save. Read the
instructions on the screen then press

ACT

to save your current

settings (or press

ESC

to cancel the save).

4

The SETTINGS SAVED message displays to confirm that your
current pump settings have been saved. Exit the menus.

Restoring pump settings

Do these steps to restore the most recent pump settings you have saved to your pump.

NOTE -

Regardless of the basal pattern you saved or the current one, Restore Settings always sets

patterns to Standard.

1

Go to the Utilities screen and select

User Settings.

Then, hold

down and

press

ACT

.

Main > Utilities > User Settings

2

The USER SETTINGS screen is displayed. Select Restore Settings
and press

ACT

.

3

A message displays, giving you the option of restoring the pump
settings that were saved on the given date and erasing the ones
currently on your pump. Read the instructions on the screen then
press

ACT

to restore the settings. (You can press

ESC

to cancel

the restore.)

ACT

Press ACT to over
write settings saved
20APR2004 at 9:41A
ESC to CANCEL

User settings have
been saved
920APR2004 at 9:43A

SETTINGS SAVED

ESC/ACT

UTILITIES MENU

Selftest

Language

User Settings

ACT

USER SETTINGS

Save Settings

Clear Settings

ACT

Restore Settings

ACT

Press ACT to restore
settings saved on
03/10/2004 at 04:15P
ESC to CANCEL