4

The SETTINGS RESTORED message displays to confirm that your
current pump settings have been replaced with the pump settings
you had saved on the given date. Exit the menus and check your
pump settings to verify the restore.

Clear settings

Take the following steps only if you want to clear your pump to factory default settings.

1

Go to the Utilities screen and select

User Settings.

Then, hold down

and press

ACT

.

Main > Utilities > User Settings

2

The USER SETTINGS screen is displayed. Select Clear Settings and press

ACT

.

3

A CONFIRM message displays, giving you the option of clearing your pump
settings or not. Select YES and press

ACT

to clear the settings. (You can

press

ESC

to cancel.)

WARNING:

Do not clear your pump settings while it is connected to your
body.

CAUTION:

Do not clear your pump settings unless directed by your
healthcare professional or a Medtronic MiniMed representative.
If you clear your pump settings, it will be necessary to
reprogram all your personal pump settings as directed by your
healthcare professional. Additionally, you will have to rewind
your pump.

SETTINGS RESTORED

User settings have been
restored.

PLEASE CHECK SETTINGS

ACT

Selftest

Language

User Settings

ACT

UTILITIES MENU

USER SETTINGS

Save Settings

Clear Settings

ACT

Restore Settings

CONFIRM

Yes

ACT

No