Alarm conditions

Alarms put the pump in “Attention” mode.

A (Alarm)

This alarm shows an “A” followed by two numbers. A-alarms cause all

insulin delivery to stop. Your pump settings are retained. If this alarm
repeats often, call the Medtronic MiniMed 24-hour Product Help Line for
assistance.

Auto off

Alerts you that no buttons were pressed during the time limit you set

for the AUTO OFF DURATION feature, and so insulin delivery has been
stopped.

Batt out limit

Occurs if the battery has been out of the pump for more than five

minutes. Verify that the pump time and date are correct. If not correct, go
to the Utilities Menu and reset the date and time.

Bolus stopped

If this alarm occurs, it is very important to check bolus history to review

how much of the bolus was actually delivered. If necessary, reprogram a
bolus with the amount not yet delivered.

Axx

9:42A

Pump reset
Settings preserved
ESC, ACT to clear

XX indicates the
alarm number

AUTO

OFF

Delivery stopped No
buttons pushed
during time limit
ESC, ACT to resume
delivery

11:17A

Batt change

too slow

ESC, ACT to clear

12:00A

OUT LIMIT

BATT

8:35A

BOLUS

STOPPED

Loose battery cap?
Pump dropped or

bumped?

Check bolus history,
Reprogram bolus

if required

ESC, ACT to clear