Chapter: 11 pump specifications, Chapter 11 pump specifications, Pump specifications – Energizer 715 User Manual

Page 165: Chapter: 11, Alarms and error messages, Alarm history, Backlight, Basal

Advertising
background image

Pump specifications 143

Chapter: 11

Pump specifications

This section provides detailed information on specifications related to your pump. The safety features of your
pump are individually listed and described.

Alarms and error messages

Indicators: audible tone (beep) or vibration (silent)

All alarms and errors show messages on the pump’s screen and provide instructions on what to do.

Unresolved alarms will escalate to siren mode for added safety.

Alarm history

The maximum records shown is 36

Backlight

Display type Liquid Crystal Display (LCD)

Time-out period is 30 seconds. The backlight will turn off when you do not push any buttons for 30

seconds or more.

Basal

Maximum patterns

delivery rate

warning limits

3 patterns (with 48 delivery
rates for each pattern)

rate increments:

0.05 units/hour

lowest possible rate:

0.05 units/hour

highest possible rate:

35.0 units/hour

factory maximum rate:

2.0 units/hour

lower limit:

less than 90 mg/dL

upper limit:

greater than 140 mg/dL