Bolus Wizard feature examples

Settings:

Carb ratio: 30 grams/unit

Insulin Sensitivity: 40 mg/dL/unit

BG Target: 90-120 mg/dL

Active Insulin Time: 6 Hours

#1: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter blood glucose.

#2: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 200.

estimate = 2 units

60g

30g/u

(food estimate)

= 2 units

200mg/dL - 120mg/dL

40mg/dL/u

= 2 + 2

= 4 units

estimate = 4 units

+

= 2 units

(correction)

60g

30g/u

(food estimate)

= 2 units