Stopping your pump, Bolus and fixed prime). see “stopping your, Suspend – Energizer 715 User Manual

Page 61: Basic programming 39

Advertising
background image

Basic programming 39

Stopping your pump

You can stop your pump with the suspend function. Suspend stops all insulin delivery including the

current basal and any bolus or prime deliveries that are in progress. While suspended, your pump will
not deliver insulin until you Resume your pump. When basal is resumed, the pump is taken out of the
Suspend mode.

The pump will beep or vibrate about every 15 minutes on the hour to remind you that it is not

delivering insulin. Example: You suspend your pump at 11:20AM. The pump will beep/vibrate at
11:30AM, 11:45AM, 12:00PM, and so on until you resume your pump (basal resumes).

NOTE -

When suspended, your pump is in Attention mode (a solid circle). When in Suspend, you can

only resume your basal or view the STATUS screen. No other functions are available.

Do these steps to suspend your pump:

NOTE -

Press

ESC

to view the STATUS screen

and verify your pump is suspended.

1

Select

Suspend

from

the MAIN MENU, and
press

ACT

.

2

SUSPEND

will flash. Press

ACT

to stop your pump.

3

The screen will show that the pump
is suspended and the time that it
stopped. After a few minutes, the
pump will default to the HOME
screen with a solid circle.

MAIN MENU

Bolus

Suspend

Basal

ACT

(flashing)

SUSPEND

Press ACT to Stop Pump

ACT

Stopped at 1:05P

SUSPEND

STATUS U100
Suspended at 01:05P
Last Bolus N 3.8U

7:58A 22 FEB

Advertising
This manual is related to the following products: