Manual prime, Disconnect, Starting on insulin 49 – Energizer 715 User Manual

Page 71

Advertising
background image

Starting on insulin 49

Manual prime

Manual prime fills the infusion set tubing with insulin before you attach it to the infusion site.

Manual prime is only available after you rewind your pump.

1

After you rewind your
pump, the PREPARING
TO PRIME screen will
appear.

WARNING: Make sure the infusion set is disconnected from your body before you press

ACT

to prime the pump. Never insert the reservoir into the pump while the

tubing is connected to your body. Doing so could result in an accidental
infusion of insulin.

2

Press and hold

ACT

to

start the prime. The
pump will beep 6 times
indicating the prime has
started.

3

While you hold

ACT

, the

pump will beep again 6 times
as the screen begins counting
the prime units being used.

ATTENTION: If you release the

ACT

button:

a.

Press

ACT

. The Manual Prime

screen appears.

b.

Press

ACT

again to continue

the prime sequence (screen
shows where you left off), or
Press

ESC

to rewind.

4

Continue to hold

ACT

until

insulin droplets form on the tip
of the infusion set needle, then
release. Be sure no bubbles are
in the tubing.

PREPARING TO PRIME

Disconnect from Body

Insert/Lock Reservoir
Hold ACT to Prime

The pump may automatically return to the HOME
screen while you install the reservoir. Press

ACT

to

return to the PREPARING TO PRIME screen.

ACT

(hold)

PREPARING TO PRIME

DISCONNECT

Hold ACT

ACT

(hold)

PRIMING – HOLD ACT

ESC if
done

U

0.3

ACT

(hold)

Advertising
This manual is related to the following products:

515