Starting on insulin 55

Fixed prime

A fixed prime fills the soft cannula with insulin and is required after the infusion set is inserted into

your body. Additionally, a fixed prime is required if you disconnect your quick release and you need to
reprime the infusion set or if you change your infusion set without changing the reservoir.

NOTE -

Prime amounts depend on the type of infusion set you are using. See your infusion set

instructions for your fixed prime amount.

Prime history

Do these steps to see a list of the delivered primes.

1

Go to the PRIME MENU.

Main > Prime

2

Select

Fixed Prime

and

press

ACT

.

3

In the FIXED PRIME screen,
enter the amount of insulin
for your type of infusion set,
then press

ACT

.

4

Once the prime begins, the PRIME
DELIVERY screen will count up the
units as they are delivered. A
“beep” will sound when priming is
complete.

1

Go to the PRIME MENU.

Main > Prime

2

Select

Prime History

and press

ACT

.

3

The PRIME HISTORY screen
will appear.

4

Scroll through the list of prime
deliveries. The “F” at the end of
the text line indicates a fixed
prime.

An “M” indicates a manual prime.
Exit the menus.

PRIME MENU

Fixed Prime

Rewind
Prime History

ACT

FIXED PRIME

ACT

U

0.

3

PRIME DELIVERY

U

0.

2

6:26P

PRIME MENU

Rewind

Prime History

Fixed Prime

ACT

PRIME HISTORY

01JAN 12:05p

03JAN 01:15p

03JAN 01:17p

22.5M

0.5F

1.2M

M

manual

prime

F

fixed

prime