Using the Bolus Wizard feature 63

More about the Bolus Wizard feature

About high or low blood glucose levels

If your blood glucose is less than 60 mg/dL (3.3 mmol/L), or

more than 250 mg/dL (13.9 mmol/L), the Bolus Wizard screen
will notify you and give you instructions. Read the instructions
and press ACT or ESC to clear the message. You can continue
programming and deliver your bolus.

About maximum delivery

The Bolus Wizard feature will not deliver more than the limit set

for your maximum bolus. If the Bolus Wizard feature calculates a
bolus amount that is larger than your max bolus limit setting, the
message, “MAX BOLUS EXCEEDED” will appear. If this happens, do
these steps:

1

In the MAX BOLUS EXCEEDED screen, press ACT to continue your
bolus programming. The estimate and maximum bolus amounts
will appear for your information. Continue to step 2.

If you do not want to continue, press ESC to cancel and the screen
will return to the ENTER BG screen.

2

In the EST: MAX screen, press ACT again to continue your bolus
programming. If desired, press ESC to cancel and the screen will
return to the ENTER BG screen.

NOTE -

Your pump will only deliver up to your maximum bolus limit setting. For example: The

Bolus Wizard feature estimate is 30 units and your max bolus limit is 25 units. When you press

ACT, your pump will only deliver 25 units, and will notify you that your bolus estimate exceeds

your max bolus.

LOW BG

DO NOT BOLUS UNTIL

BG IS NORMAL.

Treat low BG.

Monitor BG.

HIGH BG

Check for occlusion
Check ketones.

Consider insulin
injection.
Monitor BG.

ACT or ESC

- - -

MAX BOLUS

EXCEEDED

ACT

ESC

to cancel

EST: 30.0U

MAX: 25.0U

ENTER BG

mg/dL*

*or mmol/L