2 considerations for sfc/sct, Table 7, Installing and charging sfc battery pack – Honeywell ST 3000 User Manual

Page 30

Advertising
background image

3.2

Considerations for SFC/SCT

Install SFC battery
pack

If the SFC battery pack was removed for shipping and/or storage, you
will have to install the battery pack and charge the batteries before you
can operate the SFC.

The procedure in Table 7 outlines the steps for the battery pack.

Table 7

Installing and Charging SFC Battery Pack

Step Action

1

Turn SFC face down on working surface. Use metric hex wrench
(2.5 mm) to remove screws in battery compartment cover and
remove cover.

2

Insert battery pack in compartment and connect plug in compartment
to pin on battery back

Example - Battery pack installation.

Battery Pack

Hex Screws

21004

3

Replace cover and tighten hex screws

4

Connect lead from battery charger to recessed connector on left side
of SFC.

WARNING

The SFC battery charger is not intrinsically safe.

Always recharge the SFC battery pack in a nonhazardous location.
The SFC itself is an intrinsically safe device.

Continued on next page

20

ST 3000 Release 300 Installation Guide

2/05

Advertising