TROUBLESHOOTING/ERROR CODES

Condition

Possible Cause

Corrective Action

Consistent low
temperature
readings (“Lo”)

Temperature
reading is below
measurement range
(below 93.2˚F/
34.0°C).

Product Design-
Measurement range is 93.2˚F
(34.0°C) to 109.4˚F (43.0°C).
Contact your physician if you have
or suspect any medical problems.

The probe is not
positioned properly
in ear.

Make sure probe is properly
positioned against the opening of
the ear canal

The waterproof probe
is dirty.

Clean the probe with a cotton
swab moistened with alcohol.

Consistent high
temperature
readings (“Hi”)

Temperature
reading is above
measurement range
(above 109.4˚F/
43.0°C).

Product Design-
Measurement range is 93.2˚F
(34°C) to 109.4˚F (43°C). Contact
your physician if you have or
suspect any medical problems.

Consistent error
messages (“Err”)

Operating
temperature
is not within the
permissible range for
measurement.

Product Design-
Operating temperature is
60.8˚F (16°C) to 104˚F (40°C).

Low battery warning

Battery power
is low.

Replace battery (see “Inserting
and Replacing Batteries”).

Low battery warning
with “Lo” symbol

Battery power is
too low for the
thermometer to
function.

Replace battery
immediately (see “Inserting and
Replacing Batteries”).

PRODUCT SPECIFICATIONS

Model Number

Model Number TE-100

Power Source

DC 3 Volt (lithium battery CR2032)

Battery Life

1,700 Measurements

Sensor

Thermopile

Temperature Indication

In 4 digits, every 0.1˚F or ˚C

Accuracy

+/- 0.4˚F/0.2°C between 96.8˚F -
102.2˚F (36.0°C - 39.0°C)
+/- 0.5˚F/0.3°C less than 96.8˚F (36.0°C)
or greater than 102.2˚F (39.0°C)

Measurement Range

93.2˚F - 109.4˚F (34˚C - 43˚C)

Weight

0.10 lbs (with battery)

Operating Temperature

60.8˚F - 104˚F (16˚C - 40˚C)

Operating Humidity

Relative humidity up to 95%
(non-condensing)

Storage Temperature

-4˚F - 122˚F (-20˚C - 50˚C)

Storage Humidity

Relative humidity up to 95%
(non-condensing)

Contents

Ear Thermometer & Battery, Storage Case,
Instruction Manual

NOTE: These specifications are subject to change without notice. This ear
thermometer meets requirements established in ASTM Standard E1965-98.

temperature measurements, please contact HoMedics
Consumer Relations at 1-800-466-3342.