Diagnostic and error messages

Video Recorder – User Manual

251

9. Diagnostic and error messages

9.1 Diagnostic Messages

The instrument executes diagnostic routines during instrument start-up and during maintenance
procedures such as calibration. It also monitors online operation for both process faults and instrument
errors.

Error messages

Table 9-1shows messages that may appear on the instrument displays if a diagnostic condition is
detected, along with the action you should take.

Table 9-1 Diagnostic Error Messages

Message

Description

User action

INPUT FAILURE

The analog input is either open/shorted

Check input devices for open or short.

The analog input has been exposed to
electromagnetic noise.

See Section 2 for proper wiring
techniques.

The analog input is out of range

Choose a higher range.

FLOAT PT ERROR A floating point calculation error has occurred.

(divide by zero, underflow, overflow).

Check Free Form Math and Math CV
inputs for division by 0. Also check for
input values producing an underflow or
overflow error (that is, calculation is not
between -1x10

-38

and -3x10

38

or is not

between 1x10

-38

and 3x10

38

).

CHECKSUM

ERROR

An error was detected in the database of one
or more function blocks. The affected
function blocks are reset to their defaults.

Inspect entire instrument configuration
and reconfigure as needed.

ATTENTION

Be advised that the

INPUT FAILURE

diagnostic will continue to be displayed even after its cause has

disappeared. After verifying that this diagnostic’s cause has been corrected, you may clear the

INPUT

FAILURE

message from all

ON LINE

screens by performing one of the following sequences of steps.

1) Acknowledge the

INPUT FAILURE

diagnostic.

From the

ON LINE

main menu, select

ACCESS SUMMARIES

. Select

DISPLAY DIAGNOSTICS

. Use the

<Up> and <Down> buttons to highlight the

INPUT FAILURE

diagnostic(s) that you wish to acknowledge and

then press <Enter>. Once the diagnostic has been acknowledged, it will still be listed in the

DISPLAY

DIAGNOSTICS

screen but will no longer be displayed on any

ON LINE

screen.

2) Delete the

INPUT FAILURE

diagnostic.

From the

ON LINE

main menu, select

ACCESS SUMMARIES

. Select

DELETE ALL DIAGNOSTICS

. Once

this menu choice has been executed, all diagnostic messages will be deleted from the

DISPLAY

DIAGNOSTICS

screen and will no longer be displayed on any

ON LINE

screen.

During power-up, if an analog input is open or is out of range, the setpoint and process variable values
will display OFF and the loop's AUTO mode is disabled. Check connections to determine problem.
During calibration if the input reference voltages supplied by the technician are outside acceptable limits
a "CALIB FAIL" message will be generated.