Appendix b options menu features, Setting the time and date, Volume calibration – Hach 900 MAX User Manual

Page 118

Advertising
background image

Page 116
Programming Features

8990apdx_pfeatures.fm

Appendix B

Options Menu Features

1.

From the Main Menu, select

OPTIONS

.

Optional Programs

Time and Date for internal real-time clock

Calibrate the sample volume

Advanced Features

Adjust level readings of optional integral flow meter

Setting the Time and Date

1.

From the Main Menu, select

OPTIONS>TIME/DATE

2.

Adjust the real time clock and date setting, using

CHANGE AM/PM

and

CHANGE MONTH

keys.

Volume Calibration

From the Main Menu, select

OPTIONS> VOLUME CALIBRATION

.

The liquid sensor on the sampler offers unparalleled sample accuracy and
repeatability. For most site conditions, no volume calibration is necessary, and
the volume collected is accurate within 10% of the requested volume. To
improve volume accuracy to 5%, or when sample volume accuracy is
adversely affected by unusual site conditions (such as the sample source
located higher than the sampler pump), or when the liquid sensor is disabled,
then volume calibration should be performed.

Sensor Disabled
When the sensor is disabled, you must calibrate the sample volume(s). If
more than one volume is programmed (such as when using storm water or
upset sampling), you must calibrate each volume independently (see chart
below).

11:00 AM 21 - APR - 01

OPTION MENU

TIME / DATE

VOLUME

CALIBRATION

ADVANCED
OPTIONS

LEVEL

ADJUST

READY TO START

11:00 AM 21 - APR - 01

TIME / DATE

ACCEPT

CHANGE
AM / PM

_ _: _ _ AM _ _ -APR- _ _

CLEAR
ENTRY

MODE: 12-HR FORMAT

CHANGE

MONTH

USE +/- KEY TO CHANGE 12/24 HR FORMAT