Specifications, Factory installed options – Hach 900 MAX User Manual

Page 13

Advertising
background image

Specifications

Page 11

8990spc.fm

Specifications

Factory Installed Options

Intake Purge

Air purged automatically before and after each sample; duration automatically compensates
for varying intake line lengths.

Pump/Controller
Housing

High impact injection molded ABS; submersible, watertight, dust tight, corrosion & ice
resistant; NEMA 4X,6.

Internal Clock

Indicates real time and date; 0.007% time base accuracy.

Manual Sample

Initiates a sample collection independent of program in progress.

Intake Rinse

Intake line automatically rinsed with source liquid prior to each sample, from 0 to 3 rinses.

Intake Retries

Sample collection cycle automatically repeated from 0 to 3 times if sample is not obtained on
initial attempt.

Intake Tubing

¼ in. and

3

/

8

in. ID vinyl or

3

/

8

in. ID Teflon* lined polyethylene with protective outer cover.

Intake Strainers

Choice of Teflon

®

and 316 stainless construction, and all 316 stainless steel in standard size

and low profile for shallow depth applications.

pH/Temperature/ORP Meter

Control/Logging

Field selectable to log pH/temperature or ORP independent of sample operation or to control
sample collection in response to volume exceeding low/high setpoints.

pH/Temperature Sensor

Temperature compensated; impact resistant ABS plastic body

Combination electrode with porous Teflon junction

Measurement Range

0 to 14 pH, -10 to 105 °C

Operating Temperature

-18 to 80 °C (0 to 176 °F)

Dimensions

1.9 cm dia.

× 15.2 cm long (0.75 in. × 6 in.) with 1.9 cm (0.75 in.) mpt cable end

Dissolved Oxygen Meter

Control/Logging

Field selectable to log dissolved oxygen independent of sampler operation or to control
sample collection in response to volume exceeding low/high setpoints.

Measurement Method

Galvanic

Sensor

Temperature compensated; impact resistant polypropylene body

Measurement Range

0 to 20 mg/L

Resolution

0.01 mg/L

Accuracy

±3% of reading or 0.1 mg/L

Operating Temperature

0 to 50 °C (32 to 122 °F)

Dimensions

1.7 mm diameter

× 15.7 cm long (0.65 in. × 6.25 in.) with 1.9 cm (0.75 in.)

Conductivity Meter

Control/Logging

Field selectable to log conductivity independent of sampler connection or to control sample
collection in response to volume exceeding low/high setpoints.

Sensor

Temperature compensated; impact resistant polypropylene body

Measurement Range

0 to 20 mS/cm

Resolution

0.01 mS/cm or 1 mS/cm

Accuracy

±2% of reading or 0.01 mS

Operating Temperature

0 to 50 °C (32 to 122 °F)

Dimensions

1.7 cm diameter

× 15.2 cm long (0.67 in. x 6 in.) with 1.9 cm (0.75 in.) mpt cable end