Hach 900 MAX User Manual

Page 44

Advertising
background image

Page 42
Basic Programming Setup

8990startup.fm

Section 3

c.

Select either Integral or External flow meter and press

ACCEPT

.

d. Enter the flow volume between samples using the numeric keypad

and select a unit of measure using the

CHANGE UNITS

key, then press

ACCEPT

. Refer to

Table 2

for flow unit choices.

e.

Enable or Disable Timed Over-Ride using the

CHANGE CHOICE

key.

Press

ACCEPT

to continue, then enter a time period using the

numeric keypad.

Note: Select Timed Over-Ride if the flow rate drops to an unusually low value

during flow-proportional sampling and if the sample is collected once an hour
for example, even if the flow interval has not elapsed.

f.

Select Take First Sample Immediately or After First Interval?

Note: When the program is started, the first sample is taken immediately upon

pressing the start button or after the first interval has elapsed.

g. Press

ACCEPT

to continue to Sample Distribution.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

TAKE SAMPLE EVERY:

CHANGE

UNITS

1500 gal

CANCEL

CLEAR

ENTRY

ENTER: 1.00—99999999

Table 2 Sampler Pacing Flow Units

Abbreviation

Volume

gal

gallons

ltr

liters

m

3

cubic meters

af

acre-feet

cf

cubic feet

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

TIMED OVER-RIDE:

00:00 (hrs:min)

CANCEL

CLEAR

ENTRY

ENTER: 000:01 — 999:00 (hrs:min)

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

CHANGE

CHOICE

TAKE FIRST SAMPLE:

_______________

CANCEL

CHOICES: IMMEDIATE, AFTER 1

st

SAMPLE