Advanced Sampling

8990startup.fm

Section 3

The number of first flush bottles will depend on the sample volume
requirements in the NPDES permit.

b. Enter number of samples to collect using the numeric keypad.

c.

Enter the first flush sampling interval. This is the time period between
sample cycles. The first flush samples are usually collected within
thirty minutes of the storm.

Variable time intervals may be entered or the same interval may be carried
through the entire first flush. Pressing

FINAL ENTRY

carries the last displayed

interval through the remainder of the first flush sampling period.

d. Enter the first flush sample volume using the numeric keypad.

e.

Enable or Disable Program Time Limit using the

CHANGE CHOICE

key. When enabled, the Program Time Limit will stop all sampling
activity at the end of the time limit, however data logging will continue.

11:00 AM 21 - APR - 01

STORM WATER

ACCEPT

FIRST FLUSH:

NUMBER OF BOTTLES:

CANCEL

1

CLEAR

ENTRY

ENTER: 1— 4

11:00 AM 21 - APR - 01

STORM WATER

ACCEPT

FIRST FLUSH:

SAMPLER TO COLLECT:

CANCEL

6

CLEAR

ENTRY

ENTER 1 — 999

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

ACCEPT

FIRST FLUSH

INTERVAL:

CANCEL

(hrs:min)

CLEAR

ENTRY

ENTER: 000:00 — 999:00 (hrs:min)

11:00 AM 21 - APR - 01

STORM WATER

ACCEPT

FIRST FLUSH:

SAMPLE VOLUME

CANCEL

100 mL

CLEAR

ENTRY

ENTER 10 — 9999