Bump test, Calibration (excluding gasalertclip extreme) – Honeywell MicroDock II User Manual

Page 11

Advertising
background image

9

MicroDock II

Bump Test

5.

Press C exit to return to normal operation.

For more information on setting up inlets, gas types, and con-
centration levels, refer to Inlet Setup in the MicroDock II Base
Station User Manual
.

Bump Test

A bump test is performed to confirm that the detector is
responding to gas, and that the audio and visual alarms are
operational.

a

Caution

Ensure the detector is not in a low battery state prior to
performing a bump test.

To perform a bump test, complete the following:

1.

From the docking module of the applicable detector,
press C BUMP CHECK.

2.

The gas is automatically applied.

The LCD displays the results of the bump test.

If any test fails, refer to Troubleshooting in the
MicroDock II Base Station User Manual.

3.

Press C OK to return to normal operation.

4.

After 5 minutes of inactivity, the detector deactivates.

For additional information, refer to the MicroDock II
Base Station User Manual
.

Calibration

(excluding GasAlertClip Extreme)

a

Warning

BW recommends using premium grade calibration gases
and cylinders that are certified to National Standards. The
calibration gases must meet the accuracy of the detector.
All calibration cylinders must be used with demand flow
regulators and must meet the following maximum inlet
pressure specifications

• Disposable cylinders 0-1000 psig/70 bar
• Refillable cylinders 0-3000 psig/207 bar

a

Caution

Ensure the detector is not in a low battery state prior to
performing a calibration.

Advertising