4.3 Replacement Parts

Table 4-1 Replacement Parts

Description

Part Number

Complete Assembly

See Figs. 1-5 and 1-6

Cell Assembly

Cat. 4908

Complete CPVC Removal Device

31074357

Universal Head Parts

See Fig. 1-6

Platinizing Solution (3 oz. bottle)

31103011

Cell Guard Tube (0.01, 0.1, and 1.0 Constants Only)

065602

Support Tube, 1/2” NPT Sch. 80, CPVC (12” Immersion)

074344

Support Tube, 1/2” NPT Sch. 80, CPVC (6” Immersion)

074343

Junction Box

31316260

Cell Extension Leadwire (Table III other than 333)

Three conductor 18 gage cable PVC (105

°

C max.)

834059

Three conductor 18 gage cable Tefzel (150

°

C max.)

834086

9782/7082 Standard Ranges
Up to 500 ft.

Three conductor, 18 gage cable (Belden 9493)

834059

and Coax Cable (Belden 9259)

835024

Up to 1000 ft

Four conductor (3 used), 16 gage cable,

834055

Belden 9494 or equivalent and Coax Cable (Belden 9259)

835024

9782/7082 Wide Ranges

Up to 500 ft - Four Conductor, 18 gage

31834052

Up to 1000 ft - Four Conductor, 16 gage

834055

Table 4-2 Voltage and Time Limits for Platinizing Cells

Cell Constant

DC Voltage

5

10

20

25

50

1.5

----

----

----

----

----

3.0

200 sec

240 sec

----

----

----

6.0

80 sec

100 sec

180 sec

200 sec

300 sec

12.0

----

----

120 sec

150 sec