7 orp calibration, 1 one-point manual calibration, Select calibrate and confirm – Hach 6120118 User Manual

Page 30: Move the clean probe to solution and confirm, Return the probe to process

Advertising
background image

30

Operation

5.7 ORP Calibration

5.7.1 One-point Manual Calibration

The manufacturer offers a one point manual calibration for ORP. The value of the sample
used in the manual calibration may be determined by laboratory analysis or comparison
reading.

1.

From the Main Menu, select SENSOR SETUP and confirm.

2.

Select the appropriate sensor if more than one is attached and confirm.

3.

Select CALIBRATE and confirm.

4.

Select 1 POINT MANUAL CAL. Select the available Output Mode (Active, Hold, or
Transfer) from the list box and confirm.

5.

Move the clean probe to Solution and confirm.

6.

Confirm when stable. Edit the solution value using the keypad and confirm.

7.

A screen will display 1 Point Manual Complete and the slope (XX.X mV).

8.

Return the probe to process.

5.8 Concurrent Calibration of Two Sensors for pH and ORP

1.

Begin a calibration on the first sensor and continue until “Wait to Stabilize”
is displayed.

2.

Select LEAVE and confirm. The display will return to the main measurement screen.
The reading for the sensor currently being calibrated will flash.

3.

Begin the calibration for the second sensor and continue until “Wait to Stabilize”
is displayed.

4.

Select LEAVE and confirm. The display will return to the main measurement screen
and the reading for both sensors will flash. The calibration for both sensors are now
running in the background.

5.

To return to the calibration of either sensor select SENSOR SETUP from the Main
Menu and confirm. Select the appropriate sensor and confirm.

6.

The calibration in progress will be displayed. Continue with the calibration.

Advertising