SMV 3000 Transmitter User’s Manual

113

10.2

Overview,

Continued

Test Equipment
Required

Depending upon the type of calibration you choose, you may need any
of the following test equipment to accurately calibrate the transmitter:

Digital Voltmeter or milliammeter with 0.02% accuracy or better

SFC Smart Field Communicator or a PC running SCT 3000 software

Calibration-standard input source with a 0.02% accuracy

250 ohm resistor with 0.01% tolerance or better.

Using the SFC or SCT
for Calibration

Transmitter calibration can be accomplished by using either the SCT 3000
(which is recommended) or a Smart Field Communicator (SFC).

Using the SCT, calibration procedures for the SMV 3000 are available in
the on-line user manual. Step procedures for calibrating the SMV 3000
using the SFC can be found in the Smart Field Communicator Model
STS103 Operating Guide
, 34-ST-11-14.

ATTENTION

If the transmitter is digitally integrated with our TPS/TDC 3000 control
systems, you can initiate range calibration and calibration reset functions
through displays at the Universal Station. However, we still recommend
that you do a range calibration using the SCT with the transmitter removed
from service and moved to a controlled environment. Details about doing
a calibration reset through the Universal Station are given in the PM/APM
Smartline Transmitter Integration Manual,
PM12-410 which is part of the
TDC 3000 system bookset.