4 calibrating pv1 and pv2 range values – Honeywell SMV 3000 User Manual

Page 128

Advertising
background image

116

SMV 3000 Transmitter User’s Manual

1/99

10.4

Calibrating PV1 and PV2 Range Values,

Continued

Procedure, continued

Figure 28

Typical PV1 or PV2 Range Calibration Hookup

24Vdc
Power
Supply

-

+

24Vdc
Power
Supply

-

+

Low
Pressure
Port

ME T ER

SI GN A L

L

High

Pressure

Port

Absolute Pressure
Readout Device
(Smart Meter)

ST 3000 Model STA140
Absolute Pressure Transmitter
with Smart Meter

* When using a pressure source
with an absolute readout device,
such as the ST 3000 transmitter
shown, you must calibrate the LRV
for the reference absolute pressure
(or ambient atmospheric pressure)
and the URV for the absolute
pressure span.

Dead Weight Tester

or

Precision Pressure Source

(For PV1 Calibration)

Vacuum/Gauge Source

or

Precision Pressure Source *

(For PV2 Calibration)

To Low
Pressure
Port

To High

Pressure

Port

250

250

SCT

SFC

- Black

+ Red

SMV 3000

Continued on next page

Advertising