SMV 3000 Transmitter User’s Manual

125

11.4

Diagnostic Messages,

Continued

Diagnostic Messages,
continued

Table 31

Critical Status Diagnostic Message Table, Continued

SMV Status

SCT Status Message SFC Display Message TDC Status Message

Possible Cause

What to Do

7-2

Input Suspect PV3

OUTP 1 TAG ID.

SUSPCT INPUT PV3

-

PV3 Input data seems wrong.
Sensor reading is extremely
erratic.

Could be a process problem, but
it could also be a temperature
sensor or electronics module
problem.

• Cycle transmitter power OFF/ON.

• Check sensor leads for weak area

that may be ready to break or
loose connection.

3-0

Invalid Database

TAG NO.

INVALID DATABASE

INVALID DATABASE

Transmitter database was
incorrect at power-up.

Try communicating again.

Verify database configuration,
and then manually update non-
volatile memory.

7-4

NVM Fault PV3

STATUS TAG ID.

NVM FAULT PV3

NVM FAULT PV3

PV3 nonvolatile memory fault.

Replace electronics module.

8-4

Over Range PV3

STATUS TAG ID.

OVERRANGE PV3

OVERRANGE PV3

Process temperature exceeds
PV3 range.

• Check process temperature.

Reduce temperature, if required.

• Replace temperature sensor, if

needed.

9-0

PV4 (Flow) Algorithm

Parameters Invalid

STATUS TAG ID.#

ALGPARM INVALID

STATUS 9-0

Configuration for selected
equation is not complete.

Check the flow configuration using
the SCT flow compensation wizard.

3-3

PV4 in failsafe

-

STATUS 3-3

An algorithm diagnostic has
determined the flow to be invalid.

• Resolve the conditions causing

the other diagnostic message.

• Check all flow configuration

parameters.

Continued on next page