background image

130

SMV 3000 Transmitter User’s Manual

1/99

11.4

Diagnostic Messages,

Continued

Diagnostic Messages,
continued

Table 32

Non-Critical Status Diagnostic Message Table, Continued

SMV Status

SCT Status Message SFC Display Message TDC Status Message

Possible Cause

What to Do

6-4

Output Mode PV1 (DP)

STATUS TAG ID.#

OUTPUT MODE PV1

OUTPUT MODE PV1

Analog transmitter is operating as a
current source for PV1 output.

Exit Output Mode:

SCT – Press “Clear Output Mode”

button on the DP OutCal tab.

SFC – Press [

OUTPUT

] and [

CLR

]

keys.

6-5

Output Mode PV2 (SP)

STATUS TAG ID.#

OUTPUT MODE PV2

OUTPUT MODE PV2

Analog transmitter is operating as a
current source for PV2 output.

Exit Output Mode:

SCT – Press “Clear Output Mode”

button on the AP OutCal tab.

SFC – Press [

OUTPUT

] and [

CLR

]

keys.

6-6

Output Mode PV3

(Temp)

STATUS TAG ID.#

OUTPUT MODE PV3

OUTPUT MODE PV3

Analog transmitter is operating as a
current source for PV3 output.

Exit Output Mode:

SCT – Press “Clear Output Mode”

button on the TEMP OutCal
tab.

SFC – Press [

OUTPUT

] and [

CLR

]

keys.

6-7

Output Mode PV4 (Flow)

STATUS TAG ID.#

OUTPUT MODE PV4

OUTPUT MODE PV4

Analog transmitter is operating as a
current source for PV4 output.

Exit Output Mode:

SCT – Press “Clear Output Mode”

button on the FLOW OutCal
tab.

SFC – Press [

OUTPUT

] and [

CLR

]

keys.

3-7

PV4 Independent

variable out of range

-

STATUS 3-7

For R250 Laminar Flow transmitters
only. Asserted when a PV is not
within the range of a term in the
laminar Flow equation.

• Check the value of every PV

against the ranges in the
Laminar Flow equation.

• Redefine the equation, if

necessary.